INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1059
Device Problems Break (1069); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The mayfield modified skull clamp (a1059) was returned for evaluation: failure analysis: unit received with a competitor¿s plunger cap.A new plunger cap was issued to the clamp.Unit received with the lock having rotational and lateral movement and a residue buildup was present.Root cause: complaint confirmed via inspection of the unit.Unit received with the lock having movement and require replacement of worn parts as a result of normal wear and tear.Unit is beyond integra's 7 years recommended life cycle (manufactured in 2005).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
It was reported that the locking mechanism on the mayfield modified skull clamp (a1059) was broken and would not secure the patient.No patient injury or surgical delay has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12869833
MDR Text Key286760692
Report Number3004608878-2021-00650
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Device Lot Number059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Initial Date Manufacturer Received 11/03/2021
Initial Date FDA Received11/24/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-