• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 STR INSRTR THREADED SHAFT; INSTRUMENT, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 STR INSRTR THREADED SHAFT; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/12/2021
Event Type  malfunction  
Event Description
It was reported that the straight inserter handle threaded shaft thread broke off during impaction.There was no delay and no adverse consequences were reported.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).Foreign: (b)(6).The device will not be returned for analysis, due to location of device is unknown; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed with photograph received.Device history record review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR INSRTR THREADED SHAFT
Type of Device
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12888026
MDR Text Key283620068
Report Number0001825034-2021-03179
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003452
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/12/2021
Initial Date FDA Received11/29/2021
Supplement Dates Manufacturer Received12/01/2021
Supplement Dates FDA Received12/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-