Model Number 1458Q/86
Device Problems Failure to Capture (1081); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2021
Event Type  malfunction  
Event Description
It was reported that the patient presented in clinic for routine device upgrade.The left ventricular (lv) lead was placed and exhibited no capture at high threshold, failure to sense.The procedure was successfully completed with a replacement lead.The patient was stable.
 
Manufacturer Narrative
The reported events of failure to capture and sense were confirmed.A complete lead was returned.Electrical testing did not find any anomalies.Insertion testing of the lead connector into the icd device and is-4 sleeve passed.Dimensional analysis found the connector had an over-sized diameter.This may have contributed to the reported field events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12889346
MDR Text Key281411189
Report Number2017865-2021-37472
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000087820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/08/2021
Initial Date FDA Received11/30/2021
Supplement Dates Manufacturer Received02/08/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-