Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/08/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during an initial hip arthroplasty, the liner would not seat into the cup.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual inspection of returned product the post of the device and the lip both have impact marks and the od has scuffing.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|