• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-260V
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the olympus local service department on november 9, 2021, it was found that inside of the light guide lens was dirty.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
The subject device was returned to the olympus local service department.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) was informed from the olympus local service department that it was found that the subject device was leakage.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of this event could not be conclusively determined.Olympus medical systems corp.(omsc) surmised that the reported phenomenon occurred due to the following cause.- the moisture entered the subject device from leaking point and components inside the light guide lens got corroded and the corrosion product may have remained under the light guide lens as dirt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12907093
MDR Text Key285940250
Report Number8010047-2021-15311
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberTJF-260V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/09/2021
Initial Date FDA Received12/01/2021
Supplement Dates Manufacturer Received12/16/2021
Supplement Dates FDA Received01/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-