• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE 7100; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1009-9000-000
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2021
Event Type  malfunction  
Event Description
The hospital reported a malfunction resulting in loss of manual ventilation during a case.There was no patient injury.
 
Manufacturer Narrative
Ge healthcare investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Customer contact reports no patient information is available.Date of device manufacture year is 2002.The month of manufacture was unavailable at time of mdr filing.Legal manufacturer: hcs madison - (b)(4).Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The adjustable pressure limiting valve was replaced to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
MDR Report Key12908150
MDR Text Key281506271
Report Number2112667-2021-02656
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1009-9000-000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-