• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EMR; MEDICAL DEVICE DATA SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMR; MEDICAL DEVICE DATA SYSTEM Back to Search Results
Device Problem Inaccurate Delivery (2339)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Description
I am reporting a series of events related to the use of vancomycin.My institution uses precision dosing software to calculate optimal vancomycin doses.This software is not integrated or interfaced with the emr.It requires manual data entry (entering labs, medication doses, demographic information) in order to calculate doses.In a review of 86 patient courses, data entry errors occurred 17% (15).This resulted in 3% (3) of patients receiving inappropriate doses.There was no clear harm to the patient as a result of these errors.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMR
Type of Device
MEDICAL DEVICE DATA SYSTEM
MDR Report Key12908711
MDR Text Key281739999
Report NumberMW5105691
Device Sequence Number1
Product Code LRJ
Combination Product (y/n)Y
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Unknown
Type of Report Initial
Report Date 08/03/2021
3 Devices were Involved in the Event: 1   2   3  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/30/2021
Patient Sequence Number1
-
-