• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Device Problems Positioning Failure (1158); Activation Failure (3270)
Patient Problem Needle Stick/Puncture (2462)
Event Date 11/02/2021
Event Type  malfunction  
Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Based on the results of our investigation, the root cause of the complaint could not be identified.The actual sample was not returned for evaluation hence we could not provide detailed information of its actual condition.Retention samples were visually confirmed free from defects that will affect activation of safety sheath and passed evaluation for sheath activation and deactivation.Also, no irregularity was encountered during the simulation of manual sheath activation that may lead to the complaint.We have series of visual in-process inspections to detect an abnormality on the sheath that may lead to a problem during sheath activation.The molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problems and needle sticks.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.Lot history file revealed no related nonconformity or irregularity that will lead to the complaint.Prior shipment, qc conducts outgoing visual, sensory, and functional inspections to assure lots are of good quality.Therefore, we advise to follow the instructions for use (ifu) for the proper usage of the sg2 needle indicated on the unit box in which warnings to prevent needle stick, cautions, and precautions are also included.(b)(4).
 
Event Description
The user facility reported that the safety was not staying closed after the initial click.Multiple nurses have complained about it opening back up.A nurse today thought that the safety was closed however it opened and the needle stuck the nurse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON MEDI-PAK PERFORMANCE SAFETY HYPODERMIC NEEDLE
Type of Device
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna,,
RP  
Manufacturer Contact
mary o'neill
950 elkton blvd.
elkton, MD 21921
8002837866
MDR Report Key12911174
MDR Text Key281589923
Report Number3003902955-2021-00063
Device Sequence Number1
Product Code MEG
UDI-Device Identifier30612479178855
UDI-Public30612479178855
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number102-N251S
Device Lot Number210427B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-