Model Number 27001 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/15/2021 |
Event Type
malfunction
|
Event Description
|
It was reported to resmed that an astral device displayed displayed an error message (sf147) related to the main blower.There was no patient harm or serious injury reported as a result of this incident.
|
|
Manufacturer Narrative
|
The device was returned to resmed and an evaluation confirmed the complaint.Visual inspection of the main blower motor revealed contamination.The pneumatic block was replaced to address the reported complaint.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
|
|
Manufacturer Narrative
|
Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to contamination.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
|
|
Event Description
|
It was reported to resmed that an astral device displayed an error message (sf147) related to the main blower.There was no patient harm or serious injury reported as a result of this incident.
|
|