• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 100 - AMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 100 - AMER Back to Search Results
Model Number 27001
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/15/2021
Event Type  malfunction  
Event Description
It was reported to resmed that an astral device displayed displayed an error message (sf147) related to the main blower.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.Visual inspection of the main blower motor revealed contamination.The pneumatic block was replaced to address the reported complaint.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
 
Manufacturer Narrative
Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to contamination.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf147) related to the main blower.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key12911230
MDR Text Key282960890
Report Number3007573469-2021-01189
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27001
Device Catalogue Number27001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/13/2021
Distributor Facility Aware Date12/01/2021
Device Age85 MO
Date Report to Manufacturer12/13/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/01/2021
Supplement Dates Manufacturer Received12/01/2021
Supplement Dates FDA Received12/13/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2014
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-