Catalog Number A0410221G0103 |
Device Problem
Break (1069)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that a protaper gold s1 file broke.Additional information was requested for product return and further information.
|
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
Multiple unsuccessful attempts were made to obtain the patient outcome.
|
|
Manufacturer Narrative
|
The three returned protaper gold shaping files s1 21mm are actually broken in the active part (1x fatigue), or at the tip of the active part (1x fatigue; 1x mixed conditions torque + fatigue).No material defect was found during analysis of the rupture patterns.No unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1683007).Root causes are not identified.We will track this kind of event and monitor the trend.
|
|
Event Description
|
Additional information was received indicating that the broken portion of the file was removed from the patient¿s tooth.
|
|