• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLANAV ST ABLATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLANAV ST ABLATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 87069
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported that during an ablation procedure using an intellanav st ablation catheter to treat atrioventricular nodal reentry tachycardia the temperature rose higher than expected.The temperature issue occurred with every ablation attempt and no error messages halted the procedure.They exchanged the catheter and the issue was resolved.They also stated that the same set up, with different single use devices, was used previously in the day without issue.After exchanging the catheter they were able to complete the procedure without patient complications.The catheter is not expected to be returned as it was disposed of by the site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV ST ABLATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key12919919
MDR Text Key281749210
Report Number2134265-2021-15263
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/04/2024
Device Model Number87069
Device Catalogue Number87069
Device Lot Number0026922193
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/23/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-