• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGN-751
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  Injury  
Event Description
It was reported a 19mm sjm regent heart valve was chosen for procedure.During the procedure, the user reported while parachuting the valve, it was difficult to seat the valve on the annulus.The device was explanted and exchanged with a smaller, 17mm sjm regent heart valve.The event did lead to prolonged procedure, and prolonged time on the cardiopulmonary bypass.The patient is stable and is recovering.No additional information has been provided.(b)(6) 2021.(b)(6).Oed: hcp was doing avr in the patient and during the procedure hcp implanting 19mm agn valve and while parachuting the valve hcp faced some difficulty to sit the valve on the annulus so he explanted the valve and implant 17mm agn valve.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Manufacturer Narrative
An event of the valve being difficult to seat on the annulus was reported.The investigation confirmed the device met dimensional and visual specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key12920484
MDR Text Key285640149
Report Number2648612-2021-00109
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005869
UDI-Public05414734005869
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number19AGN-751
Device Catalogue Number19AGN-751
Device Lot Number7667549
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
Patient SexFemale
-
-