MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2021
Event Type  malfunction  
Manufacturer Narrative
An investigation was performed in response to a complaint of error 4163 on the aia-2000 analyzer.The device was being used for diagnosis during the complaint event.By phone, a field service engineer (fse) advised customer to wipe the tip of the cup pick-up arm with alcohol which resolved the issue.This investigation confirmed a mechanical failure due to a maintenance issue.Review of the investigation conclusions indicates that escalation of the complaint for corrective and preventive actions is not warranted.A 13-month complaint and service history review for (b)(4) from the date of (b)(6) 2020 through aware date (b)(6) 2021 was performed for similar complaints.There were no similar complaints identified during the search period.The aia-2000 operator¿s manual under appendix 4 error messages states the following: [4163] x-axis cup transfer z-axis home overrun cause: the home sensor activated improperly after movement of the x-axis cup transfer z-axis.If this occurs, the current measurement result will be flagged (mf flag) and new measurements will be suspended.Solution: contact tosoh service center or local representatives.
 
Event Description
Customer experienced an error 4163 x-axis cup transfer z-axis home overrun on the aia-2000 analyzer.An all set home completed ok.The site cleared the waste bin and chute; however, the error re-occurs after starting a run.This is a reportable event based on delay in reporting of patient results for category a analytes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key12921302
MDR Text Key287193014
Report Number3004529019-2021-00077
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-