It was reported that balloon rupture occurred.The target lesion was located in the moderately calcified superficial femoral artery.A 4.00mm/1.5cm/140cm small peripheral cutting balloon was selected for use.During the procedure, it was noted that the device was unable to inflate even when it was being inflated three times, twice for a minute at nominal pressure.Hence, a pinhole ruptured was confirmed.The device was removed without any problem and the procedure was completed with another of the same device.There were no complications reported and the patient is in good condition after the procedure.
|