ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® LIAT® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES Back to Search Results
Catalog Number 09211101190
Device Problems False Positive Result (1227); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer from thailand alleged a discrepant result for a single patient while using the cobas® sars-cov-2 & influenza a/b nucleic acid test on the cobas® liat® system.No harm is alleged.The alleged sample initially generated a positive result for sars-cov-2.The same sample was retested on a different liat analyzer which yielded a negative result for all targets.The same sample was also retested on genexpert and generated a negative result for sars-cov-2.The negative results were reported.An investigation is ongoing to evaluate the customer issue.
 
Manufacturer Narrative
A supplemental mdr will be submitted to share the conclusions of the investigation.(b)(6).(b)(4).
 
Manufacturer Narrative
Data was provided and reviewed and it was seen that the alleged run produced a positive sars result.This ct value for the positive run indicates low level of virus present and the curve appears to represent true low-level positive amplification.The data indicated the sample contains a viral target concentration at the limit of detection for the assay, so discrepancies at this level can be expected with repeat testing.The issue is sample specific and no product problem was identified.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® LIAT® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
na
branchburg NJ 08876
Manufacturer Contact
stacie-ann creighton
1080 us hwy 202 s
na
branchburg, NJ 08876
9082537112
MDR Report Key12926964
MDR Text Key283671212
Report Number2243471-2021-03839
Device Sequence Number1
Product Code QLT
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Catalogue Number09211101190
Device Lot Number10517Y
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-