The biosense webster, inc.(bwi) product analysis lab received the device on 08-oct-2021.The device evaluation was completed on 9-nov-2021.Bwi conducted a visual inspection and force evaluation of the returned device.Visual analysis of the returned sample revealed that reddish material was observed in the pebax, a hole was found on the thermocool® smart touch® sf bi-directional navigation catheter.Magnetic sensor functionality was tested on carto, and the catheter failed; errors 105 and 106 were observed.A failure analysis was performed, and the catheter was dissected on the tip area; loss of electrical continuity at the sensor was found, it was determined that the root cause was an internal failure of the sensor.The evaluation determined that the cause of pebax damage failure cannot be established.The event described reduced force accuracy was unable to be duplicated during the product investigation.However, the blood found inside the pebax area may have contributed to the high force reported.A manufacturing record evaluation was performed for the finished device (b)(4) number, and no internal actions related to the reported complaint condition were identified.As part of bwi¿s quality process all devices are manufactured, inspected, and released to approved specifications.The instructions for use contain the following warning stated in the carto 3 system manual: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.Regarding the additional finding observed, the instruction for use contains the following information: in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
It was reported that a patient underwent an atrial fibrillation (afib) procedure with a thermocool® smart touch® sf bi-directional navigation catheter (stsf) and the biosense webster inc, (bwi) product analysis lab observed a reddish material and a hole in the pebax.Initially, it was reported that the carto 3 system would lose force when going on radiofrequency (rf) with the thermocool stsf catheter.In addition, the carto 3 system displayed "reduced force accuracy" error 87 and error 85 (error message unknown by the caller).The thermocool stsf catheter was re-zeroed without resolution.The cable was replaced without resolution.The thermocool stsf catheter was replaced, and the issue was resolved.The procedure continued successfully.The carto 3 system is operating per specs and is not responsible for the product issue.No adverse patient consequences were reported.The force issue was assessed as not mdr reportable.The potential risk that it could cause or contribute to a serious injury or death to the operator or patient is remote.The issue with no force value displayed was assessed as not mdr reportable.The procedure can be completed without relying on the force data.This event is being reported because the biosense webster inc, (bwi) product analysis lab received the device for evaluation and found that there was a reddish material and a hole on the pebax.This finding was assessed as mdr reportable.Therefore, the awareness date for this reportable lab finding is (b)(6) 2021.
|