BAXTER HEALTHCARE CORPORATION AMIA AUTOMATED PD CYCLER SET; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C5479
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).Initial reporter phone number (additional): (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that there was a connection issue between the patient line of an amia automated pd cycler set and the female connector of a minicap transfer set.This occurred after use of the device for peritoneal dialysis therapy, when disconnecting.The connection issue was further described as ¿could not get disconnected" and "would not come apart at all¿.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation with the white luer adapter attached to the female connector of the transfer set.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.During functional testing, the white luer adapter was hand removed from the female connector, therefore the reported connection issue was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMIA AUTOMATED PD CYCLER SET
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12934101
MDR Text Key281755969
Report Number1416980-2021-07139
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C5479
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MINICAP TRANSFER SET
-
-