MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC EMAX 2 PLUS MOTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Model Number EMAX2PLUS
Device Problem Excessive Heating (4030)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
It was reported by the biomedical technical from france that during an unspecified surgical procedure, it was discovered that the engine of the motor device was very hot.It was not reported if there were any delays in the surgical procedure or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained, that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed.And it was determined, that the reported condition of the engine of the motor device getting very hot was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined, that the device had a component damage and displayed an e6 error code, indicating handpiece overheat warning.It was further determined, that the device failed pretest for motor thermistor assessment.The assignable root cause was determined, to be due to component failure from wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMAX 2 PLUS MOTOR
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kara ditty-bovard
4500 riverside drive
palm beach gardens, FL 33410
6103142063
MDR Report Key12953193
MDR Text Key282843199
Report Number1045834-2021-01862
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00845384002055
UDI-Public00845384002055
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K080802
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEMAX2PLUS
Device Catalogue NumberEMAX2PLUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received12/13/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-