• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION NOVALINE AK98 TUBING SET FOR HEMODIALYSIS; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION NOVALINE AK98 TUBING SET FOR HEMODIALYSIS; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number BL 11 SAP CODE 955554
Device Problem Air/Gas in Device (4062)
Patient Problem Insufficient Information (4580)
Event Date 12/01/2021
Event Type  malfunction  
Event Description
"microbubbles/small air in arterial line of novaline (ak98 machine cartridge).Not from patient vascath.Another hd staff reconnected the line to make sure not operator error.Air still seen in arterial line.Seems like air is collected/trapped in dialyzer or venous chamber and not reaching patient via venous line.Baxter product complaint line called, spoke with (b)(6).They will send sample return kit so we can send back novaline to them for review." fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVALINE AK98 TUBING SET FOR HEMODIALYSIS
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
MDR Report Key12953706
MDR Text Key281937488
Report NumberMW5105863
Device Sequence Number1
Product Code FJK
UDI-Device Identifier06948082702105
UDI-Public(01)06948082702105
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2023
Device Catalogue NumberBL 11 SAP CODE 955554
Device Lot Number2001010714
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
Patient RaceWhite, Native Hawaiian Or Other Pacific Islander
-
-