H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one bard tri-funnel replacement gastrostomy tube was returned for evaluation.Gross visual evaluation, microscopic observation and functional testing were performed for the returned product.The investigation is confirmed for the reported inflation, leak and identified fracture issue as balloon did not inflate and a longitudinal rupture was found just proximal to the distal end of the inflation lumen.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2022), g3, h6 (device, method).H11: h6 (result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|