• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC. OVERSTITCH¿ ENDOSCOPIC SUTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC. OVERSTITCH¿ ENDOSCOPIC SUTURING SYSTEM Back to Search Results
Device Problem Insufficient Information (3190)
Patient Problems Nausea (1970); Vomiting (2144); Gastrointestinal Hemorrhage (4476)
Event Type  malfunction  
Manufacturer Narrative
Additional information: the device has not been returned for analysis due to this being a literature review, and after multiple attempts to gather more information from the reporter, no additional information has been received.The investigator determined a device history record (dhr) review is not possible for this complaint, as attempts at gathering the device serial/lot number were unsuccessful.A review of the device labeling notes the following: the current overstitch¿ endoscopic suturing system (ess) instructions for use (ifu) addressed the known and potential event of "suture assembly-suture broke during procedure; vessel damage / bleeding and multiple symptoms" ess - " as follows: warning: do not introduce the device with the needle body in its open position.Warning: ensure appropriate suture slack has been created for desired suture path and pattern.Advance anchor exchange and/or manipulate endoscope to create suture slack.Anchor exchange will not install anchor onto the needle body: i.Ensure there is sufficient suture slack and the suture outside endoscope is not entangled.Ii.Ensure anchor exchange is properly positioned in alignment tube of needle driver.Warnings: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.Note: ensure suture is not tangled after removal from the cartridge.Caution: do not use when valve covers are closed as suture drag will be increased.Caution: if resistance is encountered when advancing the anchor exchange through the working channel of the endoscope, reduce the endoscope angulation until the device passes smoothly.Ensure that there is sufficient space for the needle to open.The labeling is adequate as it addresses the reported complaint.The occurrence of this reported complaint and product will be monitored as appropriate.
 
Event Description
This report is filed from an literature review where it was a review of several cases.This report is being filed for the following events: multiple symptoms and bleeding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVERSTITCH¿ ENDOSCOPIC SUTURING SYSTEM
Type of Device
ENDOSCOPIC SUTURING SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC.
1120 s. capital of texas hwy
bldg 1, ste. 300
austin TX
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, cs
CS  
Manufacturer Contact
david hooper
1120 s. captail of texas hwy
bldg 1, ste 300
austin, TX 78746
MDR Report Key12961607
MDR Text Key285242280
Report Number3006722112-2021-00116
Device Sequence Number1
Product Code OCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2021
Initial Date FDA Received12/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-