• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG SAFELIGHT FIBER OPTIC CABLE GREY 5MM X 10FT 3.05M; LIGHT SOURCE, FIBEROPTIC, ROUTINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG SAFELIGHT FIBER OPTIC CABLE GREY 5MM X 10FT 3.05M; LIGHT SOURCE, FIBEROPTIC, ROUTINE Back to Search Results
Model Number 0233050200
Device Problems Thermal Decomposition of Device (1071); Failure to Shut Off (2939)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/12/2021
Event Type  malfunction  
Event Description
It was reported that there was a potential of a thermal event.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Manufacturer Narrative
The device manufacturer date is not known.Alleged failure: "it was reported by the sales rep (b)(4) the light source remained on when it was removed from the adapter on the scope ref (b)(4) salesforce case number (b)(4)" conclusion: reported failure mode could not be reproduced but reed switch function was noted to be slightly insensitive in response to safelight adapter attachment.It's possible that these issues could cause it to get stuck in either the "on" or "off" position, leading to the reported scenario where the light source fails to turn off after the safelight adapter was removed.Probable root cause is likely due to a damaged/defective reed switch.The product was returned for investigation and the failure(s) identified is consistent with the complaint record.Failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there was a potential of a thermal event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG SAFELIGHT FIBER OPTIC CABLE GREY 5MM X 10FT 3.05M
Type of Device
LIGHT SOURCE, FIBEROPTIC, ROUTINE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
hannah stalnaker
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12965848
MDR Text Key283480622
Report Number0002936485-2021-00693
Device Sequence Number1
Product Code FCW
UDI-Device Identifier07613327051643
UDI-Public07613327051643
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0233050200
Device Catalogue Number0233050200
Device Lot Number165445
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/12/2021
Initial Date FDA Received12/08/2021
Supplement Dates Manufacturer Received11/12/2021
Supplement Dates FDA Received01/24/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-