• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH-AT SENSIS HEMO LOW; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH-AT SENSIS HEMO LOW; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 6634633
Device Problem No Visual Prompts/Feedback (4021)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Event Description
It was reported to siemens that a malfunction occurred while operating the sensis hemo low system.During an interventional procedure, the user reported that the sensis waveform went blank.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation was performed considering complaint description, cs reports, system history, and system log files.The live monitor of the sensis system went black during a scheduled interventional study.Although the system was booted up, no vital signs were displayed.As a result, the procedure was continued and finished using an alternative system.The analysis of the log file confirms an ongoing examination with a connected and monitored patient.This system consists of 2 individual pcs.One of them still displaying the vital signs.It could also be determined that the connection to the artis system was interrupted, and the user restarted the sensis system to fix the problem - without success.Siemens service was called for troubleshooting.As no system issue was found during the service intervention, the entire hardware of the signal input box (sib) was replaced as a preventive measure.After the replacement, the problem did not recur.No further details could be derived from the available information and no clear cause could be identified.A possible general error that would require corrective measures of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIS HEMO LOW
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH-AT
siemensstr. 1
forchheim, germany 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
malvern, PA 19355
6104486461
MDR Report Key12968055
MDR Text Key285825828
Report Number3004977335-2021-07974
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150493
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number6634633
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/09/2021
Supplement Dates Manufacturer Received01/17/2022
Supplement Dates FDA Received01/17/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-