It was reported that on (b)(6) 2021, a 25 mm amplatzer cribriform occluder, along with a 9 f 45/80 amplatzer trevisio intravascular delivery system, was chosen for procedure.During procedure, the left atrial disc of the device assumed a bulbous shape.The device was recaptured and repositioned, but the issue occurred again when the device was deployed a second time.The device was recaptured and removed from the patient.Another 25 mm amplatzer cribriform occluder was chosen and successfully implanted using the same delivery system.It was noted that this event was not attributed to patient anatomy.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
|
The reported event of the left atrial disc deforming into a bulbous shape not be confirmed.Three photos were received from the field, which appeared to show the distal disc in a bulbous conformation.However, the investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|