MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-010
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2021
Event Type  malfunction  
Manufacturer Narrative
An event of device deployed into cobra head was reported.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Event Description
It was reported that a 10mm amplatzer septal occluder was selected for an implant on (b)(6) 2021.During deployment the left atrial disc was deployed while pressed against the atrioventricular (av) valve rims due to the required angle of the delivery sheath.The device developed a cobra head shape.The device was retrieved into the delivery sheath and subsequent attempts to deploy the left atrial disc also revealed a cobra head shape.There was interaction with atrial structures during deployment.The device retained the original cobra shape regardless of interaction with atrial structures.The catheter was moved away from atrial structures following the first deployment.No angulation or kink noticed in the delivery system.The device was never detached from the delivery cable and was also tested outside of the patient but still retain the cobra shape outside of the patient.A 9mm amplatzer septal occluder was placed in the patient and following close evaluation was released effectively.The physician did not state any adverse events related to this portion of the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key12973139
MDR Text Key282336814
Report Number2135147-2021-00504
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010076
UDI-Public00811806010076
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-ASD-010
Device Catalogue Number9-ASD-010
Device Lot Number7234523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/16/2021
Initial Date FDA Received12/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
-
-