ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT HF; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDHFA500Q
Device Problems Failure to Deliver Shock/Stimulation (1133); Inappropriate or Unexpected Reset (2959); Compatibility Problem (2960)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Event Description
It was reported that a patient presented in clinic for a magnetic resonance imaging (mri) scan.The patient's implantable cardioverter defibrillator (icd) had gone into backup mode after the mri and defibrillation therapies were disabled.The patient was asymptomatic.The icd was successfully reprogrammed.The patient was stable throughout procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12973334
MDR Text Key282039631
Report Number2017865-2021-38935
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberP000123191
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2021
Initial Date FDA Received12/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1458Q/86 QUARTET LEAD; 2088TC/52 TENDRIL LEADS; 7122Q/65 DURATATACHY MRI LEAD
Patient Age38 YR
Patient SexFemale
-
-