We checked the returned unit and confirmed that the ccd module defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the air/water nozzle loosened, the angle wire play, the remote button cut, the lg prong top loosened, the electrical connector corroded and the lg cable connector corroded; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|