NULL PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE
|
Back to Search Results |
|
Model Number C49101320J |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, leakage of air from the product was detected.No patient injury.
|
|
Manufacturer Narrative
|
Other, other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could not be duplicated and the device passed all functional tests.Dhr review was done, no issues related to the original complaint were found.
|
|
Search Alerts/Recalls
|
|
|