A us distributor contacted zoll to report that a patient passed away in the hospital while wearing the lifevest on (b)(6) 2021.Review of the patient's download data indicates the patient received one appropriate shock and three external defibrillations on the date of passing.The device was started up at 19:40:53 on (b)(6) 2021.The patient was in an idioventricular rhythm at 40 bpm, transitioning to svt at 150 bpm with cpr/motion artifact at 03:15:20 on (b)(6) 2021.The patient's rhythm degraded to vf with cpr/motion artifact at 03:18:14.The patient received the appropriate shock at 03:18:51.The patient's rhythm at the time of the shock was vf with cpr/motion artifact.The patient's post-shock rhythm was vt at 250 bpm, self-converting to sinus bradycardia at 45 bpm with cpr/motion artifact and electrode lead fall off.The patient was in sinus bradycardia/sinus tachycardia from 30 to 140 bpm with hb and cpr/motion artifact.The patient received the first external defibrillation at 03:30:28.The patient's rhythm at the time of the shock was sinus rhythm at 80 bpm with cpr/motion artifact and electrode lead fall off.The patient's post-shock rhythm was sinus rhythm at 90 bpm.The patient received the second external defibrillation at 03:33:43.The patient's rhythm at the time of the shock and post-shock rhythm were sinus rhythm at 80 bpm with cpr/motion artifact and electrode lead fall off.The patient's rhythm degraded to vt from 150 to 160 bpm at approximately 03:36:40.The patient's rhythm then degraded to vf with cpr/motion artifact and electrode lead fall off.The patient received the third external defibrillation at 03:37:06.The patient's rhythm at the time of the shock was vf with cpr/motion artifact.The patient's post-shock rhythm was sinus tachycardia at 120 bpm with cpr/motion artifact and electrode lead fall off.Cpr/motion artifact, electrode lead fall off, and the external defibrillation prevented the lifevest from detecting the vt/vf arrhythmia.The patient was in sinus tachycardia/sinus rhythm from 120 to 60 bpm with pvc's, cpr/motion artifact and electrode lead fall off.The patient's rhythm then degraded to vt from 120 to 150 bpm with cpr/motion artifact and electrode lead fall off at 03:52:49.The patient remained in vt until the electrode belt disconnection at 03:53:49 on (b)(6) 2021.Cpr/motion artifact, electrode lead fall off, the rate of the vt arrhythmia varying below the physician-prescribed rate threshold of 150 bpm, and the electrode belt disconnection prevented the lifevest from detecting the vt arrhythmia.
|