Device evaluation of the monitor has been completed.During the incoming functional testing, a 1 hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5 hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5 hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the tactile vibration alarm, audio messaging, and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.There is no indication of a product malfunction.The electrode belt has not been recovered.The device flag data from the last download ((b)(6) 2021) does not indicate any device malfunction.The review of the data indicated that the device powered on normally and was able to acquire the patient's ecg signal on the last day of use captured in the data download.No deficiencies alleged.Device manufacture date: monitor: 05/18/2020, belt: 02/12/2021.
|
A us distributor contacted zoll to report that a patient passed away while wearing the lifevest on (b)(6) 2021.Per clinical review of the continuous ecg recording, the device was started up at 00:39:48 on (b)(6) 2021.The patient was in sinus rhythm at 90 bpm with electrode lead fall off at 13:37:58.The patient's rhythm then slowed to sinus bradycardia at 40 bpm with pvc's at approximately 13:38:37.The patient's rhythm then degraded to vt at 270 bpm with varying amplitudes, cpr/motion artifact, and electrode lead fall off at 13:38:40.The patient's rhythm then degraded to vf with varying amplitudes, cpr/motion artifact, and electrode lead fall off.The patient remained in vf until 13:57:22.Varying amplitudes, cpr/motion artifact, and electrode lead fall off prevented the lifevest from detecting the vt/vf arrhythmias.The patient's rhythm transitioned to an idioventricular rhythm from 90 to 50 bpm with electrode lead fall off at 13:57:22.The patient's rhythm degraded to asystole with crp/motion artifact and electrode lead fall off at approximately 14:37:08.The patient remained in asystole until the electrode belt disconnection at 16:40:42.
|