PHILIPS NORTH AMERICA LLC HEARTSTART XL+ DEFIBRILLATOR/MONITOR Back to Search Results
Model Number 861290
Device Problem Failure to Discharge (1169)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2021
Event Type  malfunction  
Event Description
It was reported that the device is unable to discharge.There was reportedly no patient involvement.
 
Manufacturer Narrative
The manufacturer's, authorized field service engineer (fse), evaluated the device.And confirmed, the reported problem.Upon conclusion of the evaluation.It was determined, that this was a malfunction of the power pca(printed circuit assembly) board, which was replaced.And the device was returned to full functionality.The device remains at the customer site.And no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
daniel derochers
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12981896
MDR Text Key282101004
Report Number3030677-2021-16314
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838023680
UDI-Public00884838023680
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number861290
Device Catalogue Number861290
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-