Model Number LC250 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2021 |
Event Type
malfunction
|
Event Description
|
Provider alleges, consumer alleges, the cord allegedly caught on fire.
|
|
Manufacturer Narrative
|
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
|
|
Event Description
|
Provider alleges, consumer alleges, the cord allegedly caught on fire.
|
|
Manufacturer Narrative
|
The hand control harness was damaged 12" and 18" from the connector end.There was evidence of exposed wiring but no evidence of fire or shorting.The chair was still functional despite this damage.The hand control harness was damage due to customer abuse.The owner's manual contains instruction to periodically check the hand control harnessing for damage.
|
|