MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; ELECTRIC POSITIONING CHAIR Back to Search Results
Model Number LC250
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Event Description
Provider alleges, consumer alleges, the cord allegedly caught on fire.
 
Manufacturer Narrative
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
 
Event Description
Provider alleges, consumer alleges, the cord allegedly caught on fire.
 
Manufacturer Narrative
The hand control harness was damaged 12" and 18" from the connector end.There was evidence of exposed wiring but no evidence of fire or shorting.The chair was still functional despite this damage.The hand control harness was damage due to customer abuse.The owner's manual contains instruction to periodically check the hand control harnessing for damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Type of Device
ELECTRIC POSITIONING CHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
8008008586
MDR Report Key12983291
MDR Text Key282382236
Report Number2530130-2021-00133
Device Sequence Number1
Product Code INO
UDI-Device Identifier00606509400237
UDI-Public00606509400237
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070950
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLC250
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Date Manufacturer Received01/19/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/13/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexFemale
-
-