Model Number 26921 |
Device Problem
Premature Activation (1484)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that inadvertent deployment occurred.The 60% stenosed target lesion was located in the severely tortuous and severely calcified thrombosed femoral artery.A 5 x 60 x 130 innova was selected for use.During unpacking, the stent was flushed and it was found that the device had deployed approximately 2mm.The procedure was completed with another of the same device.There were no patient complications and the patient's status was stable.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of an innova self-expanding stent system.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed that the stent is partially deployed approximately 1cm from the middle sheath.There is a kink to the outer sheath at the nosecone.Microscopic examination revealed no additional damages.The lock and pull rack are still in the manufactured location but the middle sheath is no longer sitting on top of the proximal end of the tip.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that inadvertent deployment occurred.The 60% stenosed target lesion was located in the severely tortuous and severely calcified thrombosed femoral artery.A 5 x 60 x 130 innova was selected for use.During unpacking, the stent was flushed and it was found that the device had deployed approximately 2mm.The procedure was completed with another of the same device.There were no patient complications and the patient's status was stable.
|
|
Search Alerts/Recalls
|