Model Number EL5ML
Device Problems Failure to Form Staple (2579); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4) batch #v95g6w.Investigation summary : the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Upon visual analysis of the returned sample, it was determined that the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed eight conforming clips.The event described could not be confirmed as the device performed without any difficulties noted.Although no conclusion could be reached regarding the cause of the reported event, the instructions for use do contain the following: "caution: insert the clip applier through an appropriately-sized trocar.The empty jaws will passively collapse as they are inserted through a 5 mm trocar and reopen when completely through the trocar.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Prior to loading a clip in the jaws and firing the instrument: ensure that the jaws are fully open by verifying that the line of demarcation between the jaws and the instrument shaft is past the distal end of the trocar cannula.Prior to positioning the jaws around the tubular structure or vessel, load a clip into the jaws by partially squeezing the trigger in a smooth continuous motion for approximately one-third of the total firing stroke.Position the jaws with the preloaded clip completely around the tubular structure or vessel to be ligated.The structure to be ligated should be positioned against the apex of the clip." as part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported by the sales rep that during a laparoscopic cholecystectomy the device clips were scissoring.The term ¿scissoring¿ is commonly used to describe malformed clips.After being fired, the legs of the clip did not align parallel like they are supposed to, but instead they were askew and resembled an open pair of scissors.They did not cut the tissue but they also failed to ligate the tissue.Another device was used to complete the case with no patient consequences.
 
Manufacturer Narrative
(b)(4).Date sent: 12/14/2021.D4: lot/batch number is v95g6w.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.Additional information provided: sometimes they (clips) can be very difficult to remove.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
6107428552
MDR Report Key12995545
MDR Text Key286935533
Report Number3005075853-2021-07681
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received12/14/2021
Supplement Dates Manufacturer Received11/17/2021
Supplement Dates FDA Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-