• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. CHOLEDOCHO VIDEOSCOPE; CHOLEDOCO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. CHOLEDOCHO VIDEOSCOPE; CHOLEDOCO VIDEOSCOPE Back to Search Results
Model Number CHF-V
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was inspected at (b)(4).It was confirmed that the coating of the insertion tube of the subject device was peeled off due to the deterioration.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at (b)(4), it was found the coating of the insertion tube of the subject device was peeled off.The subject device had been returned to (b)(4) for repair of the image malfunction.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The root cause of the reported phenomenon could not be identified.[consideration] based on the following information, it was presumed that the reported phenomenon caused by physical stress / chemical stress / storage environment, etc.-according to the inspection report of olympus china, insertion section was scratched and coating peeled off.-ifu states caution regarding physical stress, reprocessing method and storage environment of device.-according to the former similar case, the event had been seemingly caused by physical stress / chemical stress / storage environment, etc.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHOLEDOCHO VIDEOSCOPE
Type of Device
CHOLEDOCO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13002070
MDR Text Key282221608
Report Number8010047-2021-16090
Device Sequence Number1
Product Code FBN
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K081456
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCHF-V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/27/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/09/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-