• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CNA0T0
Device Problem Material Frayed (1262)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for analysis; the lens remains implanted.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that during an intraocular lens (iol) implant procedure, a defect on the periphery noted upon implanting lens.The surgeon did leave the lens in the patient with no patient harm.The proper loading procedure was followed.The surgeon did mention that the way see the lens in the eye was the way it delivered.She never rotated it which would put the scratch on the right side as the lens was implanted.Additional information has been requested.
 
Manufacturer Narrative
Only photo was returned.The received photo shows an implanted iol on a monitor screen, there appears to be a dark line near the edge of the optic and a dark spot in the middle of the optic.The exact location of the line/spot cannot be confirmed.A definitive determination of the reported damage cannot be made without the evaluation of the physical product.A final root cause cannot be determined based on available information.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Type of Device
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13003275
MDR Text Key282335500
Report Number9612169-2021-00293
Device Sequence Number1
Product Code HQL
UDI-Device Identifier00380652393737
UDI-Public00380652393737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P190018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2024
Device Model NumberCNA0T0
Device Lot Number25185011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2021
Initial Date FDA Received12/14/2021
Supplement Dates Manufacturer Received02/03/2022
Supplement Dates FDA Received02/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-