Model Number 9-ASD-028
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  Injury  
Event Description
It was reported that on (b)(6) 2021, a 28mm amplatzer septal occluder was implanted in a patient with a rim deficit from the posterior to the inferior side.Five hours post-implant, echo was performed and "it was confirmed that the aso became fell off from the original position" and "migrated into a pulmonary artery." the device was removed surgically and the defect was surgically closed.The patient was reported to be in stable condition.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Manufacturer Narrative
An event of device embolism in a patient with a deficit rim was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13004160
MDR Text Key284719360
Report Number2135147-2021-00538
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010212
UDI-Public00811806010212
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model Number9-ASD-028
Device Catalogue Number9-ASD-028
Device Lot Number6996670
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age37 YR
Patient SexFemale
-
-