• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 900 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 900 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Model Number TN DXH 900 HEMATOLOGY SYSTEM WITH FLOOR STAND
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Manufacturer Narrative
The field service engineer (fse) was at the customer site and inspected the instrument.To resolve the issue the fse primed sweep flow and zapped apertures to resolve the issue.The information provided reasonably suggest an instrument malfunction related to dirty apertures.Bec internal identifier - (b)(4).
 
Event Description
The customer reported getting dashes on red blood cell (rbc), rbc parameters and platelet (plt) on their dxh 900 hematology instrument.The instrument was also generating 'check for clots' errors for plt.The customer also reported erroneously high plt results for two patients.Patient data was requested but the customer was unable to provide it because they did had not separated them and did not have time to seek them out.The erroneous results were not reported out of the laboratory.There was no report of death, injury, or change to patient treatment as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 900 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196 2031
Manufacturer Contact
harry long
1000 lake hazeltine drive
m/s r590c
chaska, MN 55318
9523681224
MDR Report Key13006975
MDR Text Key282292444
Report Number1061932-2021-00177
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier15099590701574
UDI-Public(01)15099590701574(11)210524
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140911
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTN DXH 900 HEMATOLOGY SYSTEM WITH FLOOR STAND
Device Catalogue NumberC11478
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-