A us distributor contacted zoll to report that a patient passed away in the hospital while wearing the lifevest on (b)(6) 2021.It was reported that hospital staff were present and performed resuscitation efforts, including externally defibrillating the patient.Per clinical review of the continuous ecg recording, the device was started up at 11:56:36 on (b)(6) 2021.The patient was in an idioventricular rhythm at 70 bpm at approximately 12:45:12.The patient's rhythm slowed to an idioventricular rhythm at 50 bpm with cpr/motion artifact and electrode lead fall off at approximately 12:46:50.The patient's rhythm then transitioned to sinus rhythm at 80 bpm with cpr/motion artifact and electrode lead fall off at approximately 12:49:38.At approximately 12:57:18 the patient's rhythm transitioned to an idioventricular rhythm/sinus bradycardia from 60 to 30 bpm with cpr/motion artifact and electrode lead fall off.The patient remained in an idioventricular rhythm at 90 bpm with pvc's until their rhythm degraded to vt at 170 bpm with cpr/motion artifact and electrode lead fall off at approximately 13:00:59.The patient received an external defibrillation at approximately 13:01:30.The patient's rhythm at the time of the external defibrillation was vt at 130 bpm.The patient's post-shock rhythm was obscured by cpr/motion artifact and electrode lead fall off.The patient received six additional external defibrillations at 13:03:11, 13:05:06, 13:07:09, 13:10:15, 13:13:24, and 13:16:31.The patient's rhythm at the time of each external defibrillation was vf with cpr/motion artifact and electrode lead fall off.The patient's post-shock rhythm for each of the external defibrillations was obscured by cpr/motion artifact and electrode lead fall off.The patient remained in vf with cpr/motion artifact and electrode lead fall off until cpr/motion artifact and electrode lead fall off obscured the patient's rhythm at 13:20:33.Svt at 160 bpm with cpr/motion artifact and electrode lead fall off was seen at 13:20:41.The patient received an eighth external defibrillation at 13:20:44.The patient's rhythm at the time of the external defibrillation was svt at 160 bpm with cpr/motion artifact and electrode lead fall off.The patient's post-shock rhythm was obscured by cpr/motion artifact and electrode lead fall off.The patient's rhythm degraded to vf with cpr/motion artifact and electrode lead fall off at approximately 13:24:24.The patient received a ninth external defibrillation at 13:25:00.The patient's rhythm at the time of the external defibrillation was vt at 200 bpm.The patient's post-shock rhythm was obscured by cpr/motion artifact and electrode lead fall off.Cpr/motion artifact, electrode lead fall off, and the external defibrillations prevented the lifevest from detecting the vt/vf arrhythmias throughout the event.The patient was in an idioventricular rhythm from 30 to 40 bpm with cpr/motion artifact and electrode lead fall off from approximately 13:29:20 until the electrode belt disconnection at 13:35:34.The device was started up again at 21:06:37 on (b)(6) 2021.The patient's rhythm was obscured by motion artifact and electrode lead fall off until the electrode belt was disconnected at 21:08:39.
|