A us distributor contacted zoll to report that a patient passed away in a rehab facility on (b)(6) 2021.Review of the patient's download data indicates the patient received one inappropriate shock in response to oversensing of low amplitude cardiac signal and motion artifact on the date of passing.The device was started up at 14:54:38 on (b)(6) 2021.The patient was in sinus rhythm/bradycardia at 70 to 30 bpm at 18:40:02.The patient's rhythm degraded to severe bradycardia at 10 bpm with unconducted p waves.The rhythm then transitions to an idioventricular rhythm at 60 bpm before degrading back to asystole with unconducted p waves by 18:44:42.The patient received the inappropriate shock at 19:25:38.The patient's rhythm at the time of the shock and post-shock rhythm were asystole with motion artifact.The patient was last seen in asystole with cpr/motion artifact at 19:30:19.The electrode belt was disconnected at 19:31:28 on (b)(6) 2021.
|