• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TRACHEOSTOMY TUBE AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX; TRACHEOSTOMY TUBE AND TUBE CUFF Back to Search Results
Model Number 101/860/070CZ
Device Problem Deflation Problem (1149)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2021
Event Type  malfunction  
Manufacturer Narrative
One device was returned for investigation.It was leak tested and after 12 hours was found to be significantly deflated.The leak was then also confirmed by water testing.Each cuff is tested for leaks before being sent out.Due to this, the cause was traced to user error.No dhr was performed as no lot number was provided.
 
Event Description
It was reported that immediately after the customer started to use the product, deflation of the cuff was often observed.After three days in use, a severe leakage of air from the cuff was observed.A replacement was done to resolve the issue.No patient injury.No additional information available for this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX
Type of Device
TRACHEOSTOMY TUBE AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
david halverson
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key13013530
MDR Text Key282331107
Report Number3012307300-2021-12864
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076134
UDI-Public15019517076134
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/070CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2021
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-