Model Number LOW FLOW BLENDER |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire medical file identification: pc-(b)(4).Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported to vyaire medical that the low flow microblender ventilator is reading high and out of specification at a dial setting of 21%.As of this time there is no information about patient involvement associated with this reported event.
|
|
Manufacturer Narrative
|
Result of investigation: failure analysis was not able to reproduced the reported issue.Visual inspection did not show signed of physical damaged.The assembly was tested with the fio2 (fraction of inspired oxygen) set to 21% and after 1 hour no fluctuations were found.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|