• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number LOW FLOW BLENDER
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: pc-(b)(4).Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported to vyaire medical that the low flow microblender ventilator is reading high and out of specification at a dial setting of 21%.As of this time there is no information about patient involvement associated with this reported event.
 
Manufacturer Narrative
Result of investigation: failure analysis was not able to reproduced the reported issue.Visual inspection did not show signed of physical damaged.The assembly was tested with the fio2 (fraction of inspired oxygen) set to 21% and after 1 hour no fluctuations were found.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13014743
MDR Text Key282609962
Report Number2021710-2021-15160
Device Sequence Number1
Product Code BZR
UDI-Device Identifier10846446002355
UDI-Public(01)10846446002355
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLOW FLOW BLENDER
Device Catalogue Number03920A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/18/2021
Initial Date FDA Received12/15/2021
Supplement Dates Manufacturer Received04/15/2022
Supplement Dates FDA Received04/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-