MAUDE Adverse Event Report: BECTON DICKINSON UNSPECIFIED BD¿ SYRINGE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON UNSPECIFIED BD¿ SYRINGE; PISTON SYRINGE Back to Search Results
Catalog Number UNKNOWN
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 11/17/2021
Event Type  Injury  
Event Description
It was reported that 2 unspecified bd¿ syringes had the cannula break off.The following information was provided by the initial reporter: the consumer reported using the 1ml 31g ultra fine syringe and had two incidents where the needle broke off in the arm.It was stated the patient went to hospital.Date of event: unknown.Sample status: discarded.
 
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured.A catalog and lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in (b)(6) has been listed and the (b)(6) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.The customer's address is unknown.Usa has been used as a default.Medical device manufacture date: unknown.Investigation summary: exec summary no samples (including photos) were returned therefore bd was not able to duplicate or confirm the customer¿s indicated failure and the root cause is undetermined.Unable to perform complaint lot history check owing to an unknown lot number for this event.Capa/sa based on the above no additional investigation and no corrective or preventative action (capa) or situational analysis (sa) are required at this time.Dhr review: no dhr review can be carried out as the lot number is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD¿ SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13014774
MDR Text Key287676515
Report Number2243072-2021-02925
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-