MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
|
Back to Search Results |
|
Model Number BI70002000 |
Device Problems
Poor Quality Image (1408); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A manufacturer representative went to the site to test the equipment.It was initially reported that it was a known software issue with version 4.2.However, it was further noted that a problem was found with the workflow.The correct workflow was discussed, and the system was ready to be used.The imaging system then passed the system checkout and was found to be fully functional.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the images were black on the mobile viewing station (mvs). it was noted that the images still transferred from the imaging system to the navigation system with no issues, but the site could not send them to the picture archiving and communications system (pacs). no further details were available, and no patient involvement was reported.
|
|