TERUMO CORPORATION, ASHITAKA CAPIOX FX25 OXYGENATOR; OXYGENATOR, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Catalog Number CX*FX25E |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Implanted date: device was not implanted.Explanted date: device was not explanted.Phone number- requested, not provided.Health professional- requested, not provided.Occupation- requested, not provided.Pma/510(k)- k130520.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record and the product-release judgement record of the involved product code/lot# combination was conducted with no findings.(b)(4).
|
|
Event Description
|
The user facility reported the capiox device was used during the procedure.During perfusion there were noticeable high co2 values.At the end of the perfusion po2 =136mmhg at fio2 = 90%.The oxy was not part of the terumo tubing pack set.No significant delay of surgery.The patients' treatment was performed as planed and finalized.The oxy was not replaced.The procedure was completed successfully.The patient was not harmed.
|
|
Manufacturer Narrative
|
This report is being submitted as follow up no.1 to provide the device return date in section d9, update section h3, and to provide the completed investigation results.The actual sample was received for evaluation.Visual inspection upon receipt found no break or no other obvious anomaly that could lead to the decrease in the gas transfer performance.The actual sample, after having been rinsed and dried (page 3), was tested for its gas transfer performance in accordance with the factory's inspection protocol.As a result, no anomalies were revealed in the gas transfer performance of the actual sample, with the obtained values meeting the factory's control criteria.[bovine blood conditions] hb:12g/dl, temp.:37°c., ph:7.4, svo2:65%, pvco2: 45mmhg.[circulation conditions] blood flow rate: 6l/min and 4l/min, v/q=1, fio2=100%.[o2 transfer volume] @6l/min= 383 ml/min.@4l/min= 278 ml/min.[co2 removal volume] @6l/min= 338 ml/min.@4l/min= 255 ml/min.Ifu states: measure blood gases and make necessary adjustments as follows.B.Control paco2 by changing the total gas flow.To decrease paco2, increase total gas flow.To increase paco2, decrease total gas flow.Based on the provided information and investigation results, there is no definitive evidence that this event was related to a device defect or malfunction.The investigation result verified that the gas transfer performance of the actual sample after cleaned was normal meeting manufacturer control standards.In addition, no problem was found in the manufacturing-related records.Since the detailed information and the pump record were not provided, the exact cause of the reported event cannot be definitively determined based on the available information.
|
|
Manufacturer Narrative
|
This report is being submitted as follow up no.2 to provide additional information in section h10.Patient: height: 178cm, bsa: 2 review of the provided pump record revealed: circulation started at 8:53.It was stated "co2 op-feld start" at 8:54 in the event column.From this, it was confirmed that co2 gas was blown into the surgical field.Pco2 was 41 mmhg at 8:56 and increased to 88.1 mmhg at 9:11.After gas flow was increased from 5l/min to 10l/min, pco2 decreased to 49.4 mmhg.The investigation result verified that the gas transfer performance of the actual sample after cleaned was normal meeting our control standards.From the information that co2 gas was being blown into the surgical field, it was likely that blood with increased pco2 and co2 gas were suctioned into the reservoir, and blood with increased pco2 flowed into the oxygenator.However, from the condition of the actual sample, the cause of occurrence could not be clarified.
|
|
Search Alerts/Recalls
|
|
|