H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the fluency plus endovascular stent graft that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft are identified in d2 and g4.H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation.Based on the investigation of the returned delivery system, the stent graft was found partially deployed and the outer sheath was fractured.The device was flushed, and a device compatible guidewire was used.However, it was mentioned that resistance was felt when advancing the device to the target site and there was mild calcification.The fractured outer sheath indicates a higher force was applied during stent graft deployment attempts.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding the anatomy of the placement site the instructions for use states: "prior to stent graft deployment in tortuous anatomy, ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy".Regarding stent deployment, the instructions for use states "if resistance is encountered removing the delivery system, it is recommended to remove the delivery system, introducer and guidewire as a single unit".Regarding preparation of the device the instructions for use states that "prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline (.).Flushing these lumens will also facilitate stent graft deployment.Continue until saline drips from the distal end of the delivery system".Regarding accessories the instructions for use states: "a super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure; the packaging pictograms indicate an introducer size of 9f and a 0.035" guidewire.H10: b5, d4 (expiry date: 05/2024), g3, h6 (device) h11: h6 (result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|