Model Number VLV8R520
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Manufacturer Narrative
It was further informed that the device was not available for evaluation since it was discarded at hospital.Without return of the product, edwards is unable to perform a complete investigation of the reported event.Nevertheless, one picture was reviewed.Customer report of manifold broken could not be confirmed from provided picture.Picture showed a closeup on a manifold with temperature probe attached to it.Customer circled the male luer of the t-connector (with swabbable luer sites) of volumeview manifold.It was unable to determine manifold damage from provided picture.An engineering investigation has been initiated to consider any potential factors that may have contributed to this complaint.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
As reported, during use in patient with this volume view, the thermistor manifold was broken.There was not blood loss nor additional intervention required.The issue was solved replacing the catheter.There was no allegation of patient injury.Patient demographics unable to be obtained.The device was not available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLUMEVIEW
Type of Device
PROBE, THERMODILUTION
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco 00610
*  00610
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key13019712
MDR Text Key282938514
Report Number2015691-2021-06871
Device Sequence Number1
Product Code KRB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K100739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberVLV8R520
Device Catalogue NumberVLV8R520
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial