Model Number VLV8R520 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/03/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was further informed that the device was not available for evaluation since it was discarded at hospital.Without return of the product, edwards is unable to perform a complete investigation of the reported event.Nevertheless, one picture was reviewed.Customer report of manifold broken could not be confirmed from provided picture.Picture showed a closeup on a manifold with temperature probe attached to it.Customer circled the male luer of the t-connector (with swabbable luer sites) of volumeview manifold.It was unable to determine manifold damage from provided picture.An engineering investigation has been initiated to consider any potential factors that may have contributed to this complaint.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
|
Event Description
|
As reported, during use in patient with this volume view, the thermistor manifold was broken.There was not blood loss nor additional intervention required.The issue was solved replacing the catheter.There was no allegation of patient injury.Patient demographics unable to be obtained.The device was not available for evaluation.
|
|