Model Number LOW FLOW BLENDER |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
|
|
Event Description
|
The customer reported to vyaire medical blender is reading high and out of spec at a dial setting of 21%.The reporter confirmed that there is no patient involvement associated on this event.
|
|
Manufacturer Narrative
|
Result of investigation:the vyaire failure analysis laboratory received the suspect component and performed a failure investigation.The reported issue was not reproducible and no performance issues were found.A visual inspection of blender did not show signs of physical damage.The fio2 (fraction of inspired oxygen) was changed to 30%, 60%, 90% and 100% while allowing it to flow at each setting for a minimum of 20 minutes.All readings at each set point were within 2% of set value which are within the +/- 3% specification.
|
|
Search Alerts/Recalls
|