Model Number LOW FLOW BLENDER
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2021
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
 
Event Description
The customer reported to vyaire medical blender is reading high and out of spec at a dial setting of 21%.The reporter confirmed that there is no patient involvement associated on this event.
 
Manufacturer Narrative
Result of investigation:the vyaire failure analysis laboratory received the suspect component and performed a failure investigation.The reported issue was not reproducible and no performance issues were found.A visual inspection of blender did not show signs of physical damage.The fio2 (fraction of inspired oxygen) was changed to 30%, 60%, 90% and 100% while allowing it to flow at each setting for a minimum of 20 minutes.All readings at each set point were within 2% of set value which are within the +/- 3% specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13023350
MDR Text Key282389494
Report Number2021710-2021-15168
Device Sequence Number1
Product Code BZR
UDI-Device Identifier10846446002355
UDI-Public(01)10846446002355(11)20210120
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLOW FLOW BLENDER
Device Catalogue Number03920A
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-