Model Number 397002-001
Device Problems Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/03/2021
Event Type  malfunction  
Event Description
The customer, a syncardia authorized distributor, reported the companion 2 driver exhibited an emergency battery error alarm.The companion 2 driver was not supporting a patient.
 
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
 
Manufacturer Narrative
The customer-reported issue was confirmed upon review of the patient file which revealed that the driver had experienced an emergency battery error alarm and replicated during incoming functional testing when the driver exhibited an emergency battery error alarm.Additionally, during testing, the emergency battery was unable to power the driver.The root cause of the customer-reported emergency battery alarm was determined to be a malfunction of the emergency battery.The battery was found to have a permanent fault status which prohibited the battery from charging or discharging.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.Ce 5720 comp(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key13025093
MDR Text Key284633355
Report Number3003761017-2021-00238
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/06/2021
Initial Date FDA Received12/17/2021
Supplement Dates Manufacturer Received12/06/2021
Supplement Dates FDA Received07/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-