• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBBHOFF 12FR;43IN W/STYLET EN; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBBHOFF 12FR;43IN W/STYLET EN; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884711253E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2021
Event Type  malfunction  
Manufacturer Narrative
The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis is not available, and we are not able to determine the relationship between this device and the cause for this event.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.  if additional information or the sample is received, the investigation will be reopened and responded to accordingly.
 
Event Description
The customer reported that the probe presented enteral diet leakage at the y junction, where the diet/syringe connects.The tube was removed and discarded.No injuries occurred.
 
Manufacturer Narrative
The device history record was reviewed, and indicated that no abnormal process conditions were present during the manufacturing of the product that could have led to the issue reported.The device history record review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.The device was manufactured on july 14, 2020.There were no physical samples, digital images, or videos received for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported issue and determine the root cause.However, as part of continuous improvements, a corrective and preventative action has been initiated to address the reported issue.If samples are received at a later date, the complaint will be reopened, and the investigation will be updated accordingly.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 12FR;43IN W/STYLET EN
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key13025294
MDR Text Key282392672
Report Number9612030-2021-03122
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10884521582927
UDI-Public10884521582927
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711253E
Device Catalogue Number8884711253E
Device Lot Number2018902364
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2020
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-